Prospective Evaluation of the Clinical Benefits of a Novel Tattoo-less Workflow for Nonspine Bone Stereotactic Body Radiation Therapy: Integrating Surface-Guidance With Triggered Imaging Reduces Treatment Time and Eliminates the Need for Tattoos

Pract Radiat Oncol. 2024 Mar-Apr;14(2):93-102. doi: 10.1016/j.prro.2023.10.010. Epub 2023 Nov 8.

Abstract

Purpose: Oligometastatic disease has expanded the indications for nonspine bone stereotactic body radiation therapy (NSB SBRT). We investigated whether optical surface monitoring systems (OSMS) could enable tattoo-less setup and substitute for 2-dimensional/3-dimensional or cone beam computed tomography (CBCT)-based mid-imaging in NSB SBRT.

Methods and materials: OSMS was incorporated in parallel with an existing workflow using pretreatment CBCT and 2-dimensional/3-dimensional kV/kV mid-imaging beginning November 2019. The ability of OSMS to detect out-of-tolerance (>2 mm/>2°) and commanded couch shifts was analyzed. A workflow incorporating OSMS reference captures, CBCT for pretreatment verification, and OSMS/triggered imaging (TI) for intrafraction monitoring was developed for rib/sternum SBRT beginning November 2021 and all NSB SBRT beginning February 2022. Treatment time and CBCT-related radiation dose between the OSMS and the non-OSMS intrafraction monitoring group was analyzed pre- and post-OSMS/TI workflow adoption. All fractions were analyzed through statistical process control with use of an XmR chart of treatment time per quarter from February 2019 to February 2023. Special cause rules were based on Institute for Healthcare Improvement criteria.

Results: From February 2019 to February 2023, 1993 NSB SBRT fractions were delivered, including 234 rib, 109 sternum, 214 ilium, and 682 multisite. Over 20 commanded shifts, OSMS could detect 2-mm shifts to within 0.4 mm 67% of the time and 0.8 mm 95% of the time. All NSB SBRT sites showed significant reductions in treatment time, including the greatest improvement in rib total treatment (21.6-13.4 minutes; P = 1.16 × 10-17) and beam time (7.9-3.2 minutes; P = 7.32 × 10-27). Significant reductions in CBCT-related radiation were also observed for several NSB sites. These process improvements were associated with OSMS adoption.

Conclusions: Adoption of a novel NSB SBRT workflow incorporating OSMS/TI for bone intrafraction motion monitoring reduced treatment time and CBCT-related radiation exposure while also allowing for more continuous intrafraction motion monitoring for NSB SBRT. OSMS/TI enabled the transition to a tattoo-less workflow.